MCC MIIC-203 Guida Utente Pagina 62

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 77
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 61
53
Set I
2
C Bus Time-oUt in msec
Set bus time-out in milliseconds.
The iPort/USB reports a bus time-out if no inter-byte bus activity for the specified
time occurs within an I
2
C Bus message.
Command: /Unnnnn[CR] 'I
2
C Bus time-oUt (nnnn = 0 (disable)...32000 msec)
Response: * 'iPort/USB Ready
Default Setting: 10000 msec (10 seconds)
Display Firmware Version
Display firmware version
Command: /V[CR] 'Firmware Version
Response: /VCCXX.XX[CR] '(Major XX.XX Minor)
eXtended Commands
The eXtended commands are used to generate "out-of-spec" signaling. eXtended
commands cannot use the adapter’s I
2
C hardware to control the SCL and SDA lines,
as the I
2
C hardware only generates I
2
C compatible signals. The eXtended
commands use firmware to "bit-bang" the SCL and SDA lines. This firmware
cannot operate as fast as the hardware, and it can be interrupted at any time by
adapter internal interrupts. The eXtended commands run directly off the command
characters as they are received on the serial link. Speed of execution of eXtended
commands is controlled by the serial link communication rate, the execution speed
of the firmware, delays caused by execution interruptions that may occur while a
command is executing, and I
2
C Bus clock-stretching by external slave devices.
The following commands manipulate the I
2
C Clock (SCL) and data (SDA) lines.
Command: /X[S|~xx|R|r|P|0|1|?|D|d|C|c|L|A| |"]..., then Press Enter or ESCape
Enter /X followed by zero or more sub-commands, the [CR]
Response: /XCC(see commands below)[CR]
High Level Sub-Commands: